How to clean and lubricate your Telath 20:1 implant contra angle
You need to clean and lubricate your Tealth 20:1 implant contra angle after each treatment
20:1 implant contra angle is easy to be rusted by the blood and Normal Saline after implant surgery, so you need to clean and lubricate it well.
Step 1: Clean
Remove bur from handpiece after using and brush. soak the head into clean water and operate for 15 seconds.
Forward and reverse your handpiece to remove the blood and NS inside of the head.
Step 2: Drying 20:1 implant handpiece
Step 3: Lubricate it by "Lubricate helper"
Fill the bottle with lubricant oil, remove the tip of 3-way syringe, lock the device on it.
Press air button to spray. The lubricat helper with strong power can remove blood and Normal saline quickly.
Step 4: Sterilization
Place into autoclaving pouch and mark, autoclave the handpiece for 12 min. at 135℃.
Note:
● Please do not use the acid solution to scrub, inject or soak the handpiece.
● Please do not autoclave the handpiece with residual chemical substances.
● Handpiece should be packed well and place at clean and dry room without dust, sulfur and salt.
● Please do not touch the handpiece immediately after autoclaving.